Xevo TQ-S micro shows error: Desolvation temperature failed to settle - WKB24858

Article number: 24858

SYMPTOMS

  • Xevo TQ-S micro shows the error "Desolvation temperature failed to settle".
  • Source recognition: "unknown 0"

ENVIRONMENT

  • Xevo TQ-S micro

CAUSE

The probe is not properly seated.

FIX or WORKAROUND

Reseat the probe.

ADDITIONAL INFORMATION

After reseating, source recognition is 1.9 V and desolvation temperature is reached correctly.