SYMPTOMS

  • Error: Source temperature failed to settle

ENVIRONMENT

  • Xevo

CAUSE

Source temperature was inadvertently set to 20 degrees. Temperature will never get that low.

FIX or WORKAROUND

Set temperature to the normal 150 degrees.

ADDITIONAL INFORMATION

 