Low sensitivity on negative ion mode MRM transitions - WKB23945
Article number: 23945
SYMPTOMS
- Low sensitivity on negative ion mode MRM transitions in a fast switching method
- Positive ion mode transitions OK
- Negative ion mode sensitivity test passes
- If negative ions are acquired in isolation, sensitivity is as expected
ENVIRONMENT
- Xevo TQ-S micro
- Xevo TQ-S
- Xevo TQ-XS
CAUSE
Dirty ion optics and very low dwell times
FIX or WORKAROUND
- Clean ion optics: ion block, stepwave, segmented quad, MS1 quadrupole, collision cell, MS2 quadrupole.
- Set interscan and interchannel delays to automatic.
ADDITIONAL INFORMATION
As the ion optics get dirty, voltages on the optics take longer to settle when pos/neg switching. The instrument optics must be very clean to run at very short dwell times (e.g., less than 5 ms).
id23945, XEVOG2XSQT, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD