Veff changes to 1000 after updating in MassLynx - WKB239025
Article number: 239025
SYMPTOMS
- Veff can be updated for correction, but when window is opened again, the value is set to 1000
ENVIRONMENT
- Xevo G2-XS
- SYNAPT G2-Si
- MassLynx
CAUSE
This occurs when the Veff is changed in the Acquisition Settings without using the Veff Calculator.
FIX or WORKAROUND
- Open the the Acquisition Settings.
- Use the Veff Calculator to change the Veff.
- Click Update .
- DO NOT manually specify the new Veff.