Skip to main content
Waters

No MS signal, but UV signal is fine - WKB23759

Article number: 23759

SYMPTOMS

  • No MS signal.
  • UV signal is fine.
  • Create new project and methods using correct nomenclature.
  • Engineer settings are fine.
  • Found no spray coming out of probe.
  • High Pressure on system when flow directed to ESI probe

ENVIRONMENT

  • MassLynx
  • SQD
  • 2796 Alliance Bioseparations Module
  • 3100
  • Autopure FractionLynx

CAUSE

Blocked ESI probe.

FIX or WORKAROUND

  1. Clear blockage by flushing.
  2. Rebuild ESI probe using 700005795 ESI Probe Service Kit.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.