Xevo MS randomly vents - WKB234135

Article number: 234135

SYMPTOMS

  • MS randomly vents
  • Engineer dashboard shows that source randomly drops out
  • MS could pump for two hours before scroll pump cuts out
  • Xevo TQ-S micro also shows cone voltage out of range error after reboot before pump-down

ENVIRONMENT

  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof
  • Xevo TQ Absolute

CAUSE

Roughing pump overheats.

FIX or WORKAROUND

Ensure that the scroll pump runs within operating temperature limits.

ADDITIONAL INFORMATION