Skip to main content
Waters

Droplets on ESI source probe capillary tip - WKB233020

Article number: 233020

SYMPTOMS

  • No spray and no ionization on ESI source probe
  • Visible droplets from capillary probe tip

ENVIRONMENT

  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Vion IMS QTof
  • Xevo QTof
  • Xevo Tof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof

CAUSE

The ESI probe capillary position is too deep into the ESI probe tip.

 

FIX or WORKAROUND

Adjust the ESI probe capillary position so that it is visible, about 0.5 mm outside the ESI probe tip. 

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.