The percent RSD for replicate injections is very high on the Xevo TQ-S cronos - WKB232135
Article number: 232135
SYMPTOMS
- The %RSD for replicate injections is very high on the Xevo TQ-S cronos
- Sample cone is found to be very dirty
- Cleaning sample does not resolve the issue
- Sample cone cannot be cleaned completely, still shows evidence of stains
ENVIRONMENT
- Xevo TQ-S cronos
CAUSE
Sample cone was too dirty to be cleaned appropriately.
FIX or WORKAROUND
- Replace the sample cone (part 700012275).
- Replace the sample cone O-ring (700004321).
ADDITIONAL INFORMATION