The percent RSD for replicate injections is very high on the Xevo TQ-S cronos - WKB232135

Article number: 232135

SYMPTOMS

  • The %RSD for replicate injections is very high on the Xevo TQ-S cronos
  • Sample cone is found to be very dirty
  • Cleaning sample does not resolve the issue
  • Sample cone cannot be cleaned completely, still shows evidence of stains

ENVIRONMENT

  • Xevo TQ-S cronos

CAUSE

Sample cone was too dirty to be cleaned appropriately.

FIX or WORKAROUND

  1. Replace the sample cone (part 700012275).
  2. Replace the sample cone O-ring (700004321).

ADDITIONAL INFORMATION

 

id232135, XEVOTQSCRO