Fluidics leak during infusion - WKB231140
Article number: 231140
SYMPTOMS
- Fluidics leak during infusion
ENVIRONMENT
- SQD2
- Xevo TQD
- Xevo TQS
- Xevo TQ XS
- Xevo TQD
- Xevo TQ MS
- ACQUITY SQD
- ACQUITY TQD
- Xevo TQ Absolute
CAUSE
Blocked capillary causes increased back pressure during infusion.
FIX or WORKAROUND
Rebuild the probe, replacing the capillary.
Also tubing to probe or tubing between the 2 valves can be blocked.
ADDITIONAL INFORMATION
id231140, A-SQ, A-SQ, SQD, SQD2, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOTQ, XEVOTQA, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD