Skip to main content
Waters

Fluidics leak during infusion - WKB231140

Article number: 231140

SYMPTOMS

  • Fluidics leak during infusion

ENVIRONMENT

  • SQD2
  • Xevo TQD
  • Xevo TQS
  • Xevo TQ XS
  • Xevo TQD
  • Xevo TQ MS
  • ACQUITY SQD
  • ACQUITY TQD

CAUSE

Blocked capillary causes increased back pressure during infusion.

 

FIX or WORKAROUND

Rebuild the probe, replacing the capillary.

ADDITIONAL INFORMATION

 

id231140, A-SQ, A-SQ, SQD, SQD2, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD

Not able to find a solution? Click here to request help.