Skip to main content
Waters

Fluidics leak during infusion - WKB231140

Article number: 231140

SYMPTOMS

  • Fluidics leak during infusion

ENVIRONMENT

  • SQD2
  • Xevo TQD
  • Xevo TQS
  • Xevo TQ XS
  • Xevo TQD
  • Xevo TQ MS
  • ACQUITY SQD
  • ACQUITY TQD
  • Xevo TQ Absolute

CAUSE

Blocked capillary causes increased back pressure during infusion. 

FIX or WORKAROUND

Rebuild the probe, replacing the capillary.

Also tubing to probe or tubing between the 2 valves can be blocked.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.