Cannot run instrument setup on a Vion IMS QTof - WKB23075
Article number: 23075
SYMPTOMS
- IMS pressure control is not available
- Reflectron Grid voltage settle error
- Strike plate voltage settle error
- Tof Voltage settle error
- Cannot run instrument setup
ENVIRONMENT
- Vion IMS QTof
- UNIFI
CAUSE
The Vion is in full or instrument Standby.
FIX or WORKAROUND
Put the Vion into Operate.
ADDITIONAL INFORMATION