Cannot run instrument setup on a Vion IMS QTof - WKB23075
Article number: 23075
SYMPTOMS
- IMS pressure control is not available
- Reflectron Grid voltage settle error
- Strike plate voltage settle error
- Tof Voltage settle error
- Cannot run instrument setup
ENVIRONMENT
- Vion IMS QTof
- UNIFI
CAUSE
The Vion is in full or instrument Standby.
FIX or WORKAROUND
Put the Vion into Operate.
ADDITIONAL INFORMATION
id23075, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC, VION