Source pressure test fails; high exhaust warning on QDa or Xevo instruments- WKB22380
Article number: 22380
SYMPTOMS
- Source pressure test fails
- High exhaust warning on QDa instrument
ENVIRONMENT
- Xevo TQD
- Xevo TQ MS
- Xevo TQ-S
- Xevo TQ-XS
- Xevo G2-Tof
- Xevo G2-QTof
- Xevo G2-S
- Xevo G2-XS
- SYNAPT G2 MS
- SYNAPT G2 HDMS
- SYNAPT G2-S
- SYNAPT G2-Si
- QDa
CAUSE
Solvent is collecting at a low point in the exhaust line between the instrument and the trap bottle.
FIX or WORKAROUND
- Drain the solvent from the exhaust tubing.
- Move the trap bottle so that there is no low point between the instrument's exhaust port and the trap bottle.
ADDITIONAL INFORMATION
id22380, A-SQ, QDA, SQD2, SYNAPT, SYNAPTMS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, TQD, TQDIVD, XEVOG2QTOF, XEVOG2TOF, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD