Source pressure test fails; high exhaust warning on QDa or Xevo instruments- WKB22380

Article number: 22380

SYMPTOMS

  • Source pressure test fails
  • High exhaust warning on QDa instrument

ENVIRONMENT

  • Xevo TQD
  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQ-XS
  • Xevo G2-Tof
  • Xevo G2-QTof
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2 MS
  • SYNAPT G2 HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • QDa

CAUSE

Solvent is collecting at a low point in the exhaust line between the instrument and the trap bottle.

FIX or WORKAROUND

  1. Drain the solvent from the exhaust tubing.
  2. Move the trap bottle so that there is no low point between the instrument's exhaust port and the trap bottle.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.