Skip to main content
Waters

Automatic calibration fails for NaICs on a SYNAPT - WKB222124

Article number: 222124

SYMPTOMS

  • Automatic calibration gives variable results when run on the SYNAPT
  • Sometimes passes with <1 ppm error, other times a >200 ppm error is observed

ENVIRONMENT

  • MassLynx
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • IntelliStart

CAUSE

An automatic calibration on NaICs is know to sometimes pick the wrong peak near 700 m/z, and it is noted in the reference file for the compound.

FIX or WORKAROUND

Run an assisted calibration. If the wrong peak is picked, ensure that it is manually changed to the correct peak. The error for the calibration should be <1 ppm.

ADDITIONAL INFORMATION

 

id222124, MALDIG2SI, MLYNX, MLYNXV41, SUPMM, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML

Not able to find a solution? Click here to request help.