Cannot acquire data when using Dynamic Range Extended function with Xevo QToF - WKB207112
Article number: 207112
SYMPTOMS
- Can acquire data with functions that start from 0 min
- Cannot acquire data with functions that start from other than 0 min
ENVIRONMENT
- Xevo G2-XS
- Xevo G3 QToF
- Dynamic Range Extended
CAUSE
Data Format was set to Continuum and the Dynamic Range Extended function was used.
FIX or WORKAROUND
When using the Dynamic Range Extended function, set Data Format to Centroid.
If you need to acquire the Data Format in Continuum, use the MS method with Normal selected instead of Extended.