Skip to main content
Waters

No reference probe signal - WKB207056

Article number: 207056

SYMPTOMS

  • No signal over the reference probe
  • A "clack, clack" sound is heard coming from the reference syringe pump
  • Disconnecting the black tubing (fluidics-reference probe) from the source side does not stop the noise
  • Disconnecting the tubing from the fluidics makes the noise disappear

ENVIRONMENT

  • Xevo QTof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS
  • SELECT SERIES Cyclic IMS

CAUSE

The black tubing between the reference fluidics and the reference entrance in the enclosure is clogged, causing overpressure in the syringe pump.

FIX or WORKAROUND

Replace the black tubing or cut both ends by 0.5cm

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.