"Turbo Speed Not Sufficient for Operation" error message with ACQUITY QDa Mass Detector - WKB206720
Article number: 206720
SYMPTOMS
- "Turbo Speed Not Sufficient for Operation" error message with ACQUITY QDa Mass Detector
- Rear fans are noisy and operating above normal speed
ENVIRONMENT
- ACQUITY QDa Mass Detector
CAUSE
The aperture disk/seal is not seated correctly.
FIX or WORKAROUND
Re-seat the aperture disk/seal.
ADDITIONAL INFORMATION
id206720, QDA
What are the part numbers for the source aperture carrier, aperture disk, and sample cone for an ACQUITY QDa? - WKB52732