MassLynx begins acquiring data without waiting for inlet start when using APGC - WKB206621
Article number: 206621
SYMPTOMS
- No injection on APGC option fitted to the Xevo MS
- The GC inlet shows "waiting for preprun", although the GC inlet is in a "Ready" state
- MassLynx immediately starts acquiring data. without waiting for an injection start signal from the GC
ENVIRONMENT
- MassLynx 4.2
- APGC
- Xevo quadrupole MS
- Xevo TOF MS
CAUSE
Injection volume is set to zero (0) in the sample list.
FIX or WORKAROUND
In the MassLynx sample list, set the injection volume to an appropriate value other than zero.
ADDITIONAL INFORMATION