Oil leakage on Edwards nRV14i rotary pump - WKB206322

Article number: 206322

SYMPTOMS

  • Oil leaks on the top of the pump and forms a pool
  • pump.png

ENVIRONMENT

  • Xevo TQ-S cronos
  • ACQUITY RDa Detector

CAUSE

The connection between oil mist filter and pump is not tight enough with the plastic clamp.

FIX or WORKAROUND

Replace the plastic clamp with the metal clamp and tighten.

ADDITIONAL INFORMATION