Negative calibration is failing on Synapt G2-S and Synapt G2 - WKB203615
Article number: 203615
SYMPTOMS
- negative calibration routine is failing
- detector setup not run in 8 months
ENVIRONMENT
- MassLynx
- Xevo G2-S
- Xevo G2-XS
- Synapt G2-S
- SYNAPT G2
- SYNAPT G2 MS/HDMS
CAUSE
Not optimized detector voltage
FIX or WORKAROUND
- Run negative mode detector setup with leucine encaphalin standard.
- Once setup passes, rerun calibration routine.
ADDITIONAL INFORMATION
id203615, MLYNX, MLYNXV41, SUPMM, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIMS, SYNG2SMS, SYNG2SMSML