Skip to main content
Waters

Negative calibration is failing on Synapt G2-S and Synapt G2 - WKB203615

Article number: 203615

SYMPTOMS

  • negative calibration routine is failing
  • detector setup not run in 8 months

ENVIRONMENT

  • MassLynx
  • Synapt G2-S
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS

CAUSE

Not optimized detector voltage

FIX or WORKAROUND

  1. Run negative mode detector setup with leucine encaphalin standard.
  2. Once setup passes, rerun calibration routine.

ADDITIONAL INFORMATION

 

id203615, MLYNX, MLYNXV41, SUPMM, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIMS, SYNG2SMS, SYNG2SMSML

Not able to find a solution? Click here to request help.