Calibration file from IntelliStart does not remain linked to the tune file in MassLynx for a Xevo TQ-S micro - WKB202615

Article number: 202615

SYMPTOMS

  • Error message: No calibration file was found for this analysis when trying to run Sample Tune and Develop Method from IntelliStart
  • The existing calibration was cleared from the Calibrate > Instrument calibration dialog on the tune page

ENVIRONMENT

  • MassLynx 4.2
  • IntelliStart
  • Xevo TQ-S micro

CAUSE

Incorrect Type of Analysis selected in the IntelliStart Setup Parameters

FIX or WORKAROUND

  1. In the MS Console, go to the IntelliStart page.
  2. Go to Configure > IntelliStart Setup Parameters.
  3. Select the appropriate Type of Analysis.

ADDITIONAL INFORMATION

 

id202615, MLYNX, MLYNXV41, SUPMM, XEVOTQSMIC