loss of sensitivity Xevo MS - WKB200902

Article number: 200902

SYMPTOMS

  • poor sensitivity
  • glowing discharge on capillary tip. 
  • residue at base of source enclosure

ENVIRONMENT

  • Xevo TQD
  • Xevo TQ-S Micro
  • Masslynx

CAUSE

Capillary voltage is being set too high. 

FIX or WORKAROUND

  1. The capillary voltage was tuned to a more optimized voltage for the LC flowrate ( <2Kv)

ADDITIONAL INFORMATION