No capillary voltage; desolvation gas cannot be turned on because of active probe interlock on LCT Premier XE - WKB196604

Article number: 196604

SYMPTOMS

  • No capillary voltage 
  • Capillary voltage readback is 0
  • Desolvation (API) gas cannot be turned on
  • Probe interlock active message at the bottom of the tune page

ENVIRONMENT

  • LCT Premier
  • LCT Premier XE

CAUSE

The LockSpray baffle motor panel is not fully seated, causing the interlock to be active.

FIX or WORKAROUND

Tighten or reseat the LockSpray baffle motor panel.

ADDITIONAL INFORMATION

 

id196604, LCTPREMIER, LCTPREMXE