No capillary voltage; desolvation gas cannot be turned on because of active probe interlock on LCT Premier XE - WKB196604
Article number: 196604
SYMPTOMS
- No capillary voltage
- Capillary voltage readback is 0
- Desolvation (API) gas cannot be turned on
- Probe interlock active message at the bottom of the tune page
ENVIRONMENT
- LCT Premier
- LCT Premier XE
CAUSE
The LockSpray baffle motor panel is not fully seated, causing the interlock to be active.
FIX or WORKAROUND
Tighten or reseat the LockSpray baffle motor panel.
ADDITIONAL INFORMATION
id196604, LCTPREMIER, LCTPREMXE