Source leaks at the union connecting to LC on sample probe - WKB196536
Article number: 196536
SYMPTOMS
- Severe leaking from the union on ESI sample probe when the LC connection is removed
ENVIRONMENT
- QTof Premier
CAUSE
The sample capillary is pushed out from the probe.
FIX or WORKAROUND
Rebuild the sample probe.
ADDITIONAL INFORMATION
id196536, Electrospray, QTOF-PREM