Xevo TQ-S "Low desolvation gas flow" error - WKB193588
Article number: 193588
SYMPTOMS
- Low desolvation error in console logs
- Input N2 Pressure at regulator is OK, ~100 psi
- Source pressure test successful
- Cone gas flow readback is normal (150 L/h) and nebulizer readback is normal (6.9 bar)
- Issue persists after a hard reboot of the MS using the rear switches
ENVIRONMENT
- Xevo TQ-S
- MassLynx
CAUSE
Frozen or locked desolvation setting on the tune page.
FIX or WORKAROUND
- Change the set desolvation gas flow to 1200 L/h, and then return to the normal set flow rate.
- The N2 readback should now work properly.
ADDITIONAL INFORMATION
MassLynx SCN986 with Full Security
id193588, SCN 986, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD