Skip to main content
Waters

Source Pressure Test failure, leak warning - WKB18801

Article number: 18801

SYMPTOMS

  • Source Pressure Test leak warning
  • SPT leak

ENVIRONMENT

  • SQ Detector 2
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • Xevo Qtof
  • Xevo Tof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS

CAUSE

Source enclosure not sealing correctly. This can be a result of debris preventing the seals from making good contact or failure of the seals themselves.

FIX or WORKAROUND

Open the source enclosure door and clean the seals in situ. Closure the source door and run the Source Pressure Test. If the issue persists, replace the source enclosure seals and repeat

Part number for Outer API Source Service Kit (# 700004670)

ADDITIONAL INFORMATION

id18801, MALDIG2SI, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, VION, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVTQSMIVD

Not able to find a solution? Click here to request help.