Source Pressure Test failure, leak warning - WKB18801

Article number: 18801

SYMPTOMS

  • Source Pressure Test leak warning
  • SPT leak

ENVIRONMENT

  • SQ Detector 2
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • Xevo Qtof
  • Xevo Tof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS

CAUSE

Source enclosure not sealing correctly. This can be a result of debris preventing the seals from making good contact or failure of the seals themselves or an issue with the door latch mechanism

FIX or WORKAROUND

Open the source enclosure door and clean the seals in situ. Closure the source door and run the Source Pressure Test. If the issue persists, replace the source enclosure seals and repeat

Part number for Outer API Source Service Kit (# 700004670)

If the issue continues to persist consider replacing the source enclosure

ADDITIONAL INFORMATION