Tune page is stuck at 200000 Da for the GCT Premier - WKB18055

Article number: 18055

SYMPTOMS

  • Mass range stuck at 200000 on tune page
  • D8 on GCT Control PCB flashes once when the TDC settings are refreshed

ENVIRONMENT

  • GCT Premier

CAUSE

No trigger signal to the TDC card because sync cable from the pusher PCB is broken.

FIX or WORKAROUND

Replace the pusher unit (part 4221001DC1-S).

ADDITIONAL INFORMATION