Skip to main content
Waters

QDa is not communicating in MassLynx; instrument not present - WKB17931

Article number: 17931

SYMPTOMS

  • "Instrument not present" warning in MassLynx
  • No control over the QDa

ENVIRONMENT

  • MassLynx
  • ACQUITY QDa Mass Detector

CAUSE

Improper or too rapid boot sequence

FIX or WORKAROUND

  1. Close out of MassLynx.
  2. Power-off the QDa.
  3. Reboot the PC and log in to Windows, but do not start MassLynx.
    • Allow the loading of all required services and processes to complete.
    • Power on Acquity with Column Manager or Sample Manager (whichever has LAN cables).
    • Or power-on all other components, starting with the 515/PCM or makeup pump.
  4. Power-on the QDa and leave it for 15 minutes to go through the boot sequence.
    • Allow time for the detector to perform startup checks properly and for the MS to pump down.
  5. Start MassLynx and verify that the QDa connects (should say "Not Scanning").

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.