The oil trap is full on the Vacuubrand pump connected to the QDa - WKB1785

Article number: 1785

SYMPTOMS

  • The oil trap on the external roughing pump (perfomance only) is full.

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

The pump doesn't return oil to the reservoir, the operator must do it manually.

FIX or WORKAROUND

Follow the procedure in the user guide for replacing the oil. ACQUITY QDa Detector Overview and Maintenance Guide