How to make a new lock mass profile - WKB17669

Article number: 17669

OBJECTIVE or GOAL

Make a new lock mass profile.

ENVIRONMENT

  • Xevo G2-S
  • MassLynx
  • IntelliStart

PROCEDURE

  1. Select Configuration mode in IntelliStart.
  2. From the Configure pull-down menu, choose Lock Mass Editor.
  3. In the window that appears, click File, and then click New.
  4. Specify a profile name, and then complete the rest of the window with appropriate values.
  5. Click OK to complete.
  6. Ensure that your new profile appears in the list in the Lock Mass Editor.