Which tune page is required when using instrument setup for a Xevo G2-S, Xevo G2-S, Xevo G2-XS, SYNAPT G2, SYNAPT G2-S or SYNAPT G2-Si? - WKB16807

Article number: 16807

ENVIRONMENT

  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si

ANSWER

When doing calibration and detector setup – Use a tune page with low desolvation temperature and low desolvation gas flow (for example, 200 °C and 400 L/hr).

When doing lockspray setup – Use the tune page that will be used for the LC-MS runs (that is, high desolvation temperature and gas flow). Be sure to have the LC pump flowing.