Sodium adducts in much higher abundance than the protonated ion on SYNAPT G2-Si - WKB16747
Article number: 16747
SYMPTOMS
- When infusing sample, see many more sodiated [M+Na]+ or [M+23]+ ions than protonated [M+H]+ or [M+1]+ ions
- Peak of interest is 22 Da higher than expected for singly charged ions
ENVIRONMENT
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- Xevo TQD
- Xevo TQ-MS
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
- Xevo QTof
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
CAUSE
High sodium, salt, or soap contamination in the source enclosure area
FIX or WORKAROUND
- Wipe down the source enclosure with methanol.
- Infuse 90/10 water/methanol + 1% formic acid.
- Clean the cone and baffle (if there is one).
ADDITIONAL INFORMATION
id16747, MALDIG2SI, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, XEVOTQD, XEVOTQDIVD