QTof Premier's source temperature will not go above 30 °C - WKB16429
Article number: 16429
SYMPTOMS
- The readback for the source temperature will not go above 30 °C, even though a higher source temperature is specified on the tune page
- The cartridge heater assembly (700001201) was replaced, but still no change
- "Source Fuse" LED is lit on the Source PCB.
ENVIRONMENT
- QTof Premier
- MassLynx 4.1
CAUSE
Fuse F1 on the source connector PCB has blown.
FIX or WORKAROUND
- Put system into pump override.
- Turn off electronics breaker.
- Remove fuse F1 and replace it.
- Turn on electronics breaker and test system.
ADDITIONAL INFORMATION
Refer to the service note "FIRMWARE: Source heater fuse blowing - new Source PCB firmware (Issue J)". It is very important to know what firmware and fuse are being used. An incorrect pairing could cause source meltdown.
id16429, MLYNX, MLYNXV41, QTOF-PREM, SUPMM