How to start the vacuum system on a Xevo G2-XS ToF - WKB16353

Article number: 16353

OBJECTIVE or GOAL

Start the instrument vacuum system.

ENVIRONMENT

  • Xevo G2 ToF
  • Xevo G2-S ToF
  • Xevo G2-XS ToF
  • MassLynx 4.1

PROCEDURE

  1. Open the MS Console.
  2. Click the Operate button.
  3. The backing pumps will come on automatically and, when backing pressure achieves the trip level, the turbos will speed up.

ADDITIONAL INFORMATION

 

id16353, MLYNX, MLYNXV41, SUPMM, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF