Vion fails mass calibration - WKB16314
Article number: 16314
SYMPTOMS
- Mass calibration fails
- Fluidics were purged multiple times
- The beam intensity is stable
- No ions or minimal ions are observed above 1,000 when infusing major mix in positive ion mode
ENVIRONMENT
- Vion
- UNIFI
CAUSE
The major mix calibration solution has degraded.
FIX or WORKAROUND
- Prepare fresh major mix (186008113 - Major Mix Tof/IMS Calibration Kit).
ADDITIONAL INFORMATION
See the attached document (715005131 MajorMix Prep.pdf) for instructions on major-mix preparation.
id16314, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC, VION