How to set up post-column addition on a Xevo or SYNAPT G2 mass spectrometer - WKB15356

Article number: 15356

OBJECTIVE or GOAL

Set up post-column addition using the divert valve.

ENVIRONMENT

  • SQD2
  • Xevo TQD
  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • Xevo QTof/Tof
  • Xevo G2 QTof/Tof
  • Xevo G2-S QTof/Tof
  • Xevo G2-XS QTof/Tof
  • SYNAPT G2 QTof/Tof
  • SYNAPT G2-S QTof/Tof
  • SYNAPT G2-Si QTof/Tof

PROCEDURE

  • Connect the post-column addition solution via one of the fluidics lines. You can switch the line that goes to a fluidics bottle at the valve to run to a solvent bottle if needed.
  • Set up the MS method events as follows, adjusting the flow rate, timing, and so on to the desired parameters. For the quads, the method event is found in the Options drop-down menu > Method events. For the Tofs, there is a button in the MS method called Method Events. method events.2PNG.PNG
  • On a quad, if not diverting to waste, it should look like:

Quad_method_events no divert.PNG

  • On a Quad, if diverting to waste and then going to combined flow, it should look like:

Quad_method_events divert.PNG

  • On a QTof/Tof, if not using the divert valve to go to waste, it should look like:

Tof method events no divert.PNG

  • On a QTof/Tof, if using the divert valve to go to waste, it should look like:

Tof method events divert.PNG

ADDITIONAL INFORMATION

 