MS console shows "fluidics operation stopped" error when using APGC source- WKB1498
Article number: 1498
SYMPTOMS
- MS console shows a blinking red System Status and the message "fluidics operation stopped"
ENVIRONMENT
- APGC
- MassLynx 4.1 SCN932
- MassLynx 4.2 SCN1007
- SYNAPT G2-Si
- Xevo TQ-XS
CAUSE
IntelliStart configurations/definitions for the Fluidics System Error, Reference Fluidics System error, and Fluidics Operation Stopped choices are enabled.
FIX or WORKAROUND
- Go to the MS Console.
- Select SYNAPT G2-Si / Xevo TQ-XS from the Instrument tree.
- Go to the "Troubleshoot" menu.
- Select "Service Mode" and type the password 1147.
- Go back to the SYNAPT G2-Si / Xevo TQ-XS menu.
- Go to the "Configure" menu.
- Select "IntelliStart Configurations".
- Clear the selections of the,
-Fluidics System Error
-Reference Fluidics System Error
-Fluidics Operation Stopped
ADDITIONAL INFORMATION
-These IntelliStart configurations are not required for APGC mode.
-Tune page's Fluidics tab will stay at the status of, "Stopped - Select purge to recover".
id1498, APGC, MLYNX, MLYNXV41, SCN 1007, SCN 932, SUPMM