Column manager not shown in MassLynx inlet editor - WKB14149

Article number: 14149

SYMPTOMS

  • Column manager is missing from MassLynx Inlet editor

ENVIRONMENT

  • MassLynx
  • ACQUITY Column Manager

CAUSE

The column manager is not selected in the instrument configuration.

FIX or WORKAROUND

  1. Start MassLynx and open the Inlet Editor.
  2. Select Tools.
  3. Select Instrument Configuration.
  4. Click Configure.
  5. Click Next until the Add Detectors page appears.
  6. In the left-hand list, locate Waters ACQUITY CM and use the arrow button to move it to the right-hand list.
  7. Click NEXT twice.
  8. Click Finish.
  9. Click Finish.
  10. Do not run setup.
  11. The CM is now added.