Incorrect isotopic pattern for mass spectral peaks on SYNAPT G2-S - WKB14114
Article number: 14114
SYMPTOMS
- The isotope pattern is off for intense peaks on the SYNAPT G2-S
ENVIRONMENT
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
CAUSE
Detector saturation is affecting the isotope pattern.
FIX or WORKAROUND
Run samples using the DRE lens option in the MS method to get the proper isotope pattern.
ADDITIONAL INFORMATION