What is an acceptable desolvation gas readback for the Xevo TQD? - WKB13506
Article number: 13506
ENVIRONMENT
- Xevo TQD
ANSWER
Up to 10 under the set point is acceptable flow for the desolvation gas. Example: for a set point of 800, a readback of 790 is okay.
ADDITIONAL INFORMATION
If the nitrogen supply is in question, run the source pressure test by opening and closing the source enclosure. If it passes, the nitrogen supply is fine.