Xevo TQD shows no chromatographic peaks for any samples - WKB13446
Article number: 13446
SYMPTOMS
- No peaks when injecting or infusing
- Vials pierced
- No error messages
- Tune page readbacks OK
- System pressure normal
- No leaks
- Rebooted system
ENVIRONMENT
- Xevo TQD
- ACQUITY TQD system
- TQ Detector
- Xevo TQ MS
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
- MassLynx
CAUSE
The isolation valve is closed.
FIX or WORKAROUND
Open the isolation valve. On the Xevo TQD, the isolation valve is open when the arrow is horizontal.
ADDITIONAL INFORMATION
id13446, MLYNX, MLYNXV41, SUPMM, TQD, TQDIVD, XEVOTQD, XEVOTQDIVD