Xevo TQD shows no chromatographic peaks for any samples - WKB13446
Article number: 13446
SYMPTOMS
- No peaks when injecting or infusing
- Vials pierced
- No error messages
- Tune page readbacks OK
- System pressure normal
- No leaks
- Rebooted system
ENVIRONMENT
- Xevo TQD
- ACQUITY TQD system
- TQ Detector
- Xevo TQ MS
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
- MassLynx
CAUSE
The isolation valve is closed.
FIX or WORKAROUND
Open the isolation valve. On the Xevo TQD, the isolation valve is open when the arrow is horizontal.
ADDITIONAL INFORMATION