Skip to main content
Waters

Xevo TQD shows no chromatographic peaks for any samples - WKB13446

Article number: 13446

SYMPTOMS

  • No peaks when injecting or infusing
  • Vials pierced
  • No error messages
  • Tune page readbacks OK
  • System pressure normal
  • No leaks
  • Rebooted system

ENVIRONMENT

  • Xevo TQD
  • ACQUITY TQD system
  • TQ Detector
  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • MassLynx

CAUSE

The isolation valve is closed.

FIX or WORKAROUND

Open the isolation valve. On the Xevo TQD, the isolation valve is open when the arrow is horizontal.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.