Communication error between the ACQUITY PDA detector and the ACQUITY MS Console - WKB13238
Article number: 13238
SYMPTOMS
- PDA Detector Communication Failure error in the ACQUITY MS Console
- Rebooting PDA Detector does not resolve
- Found: Status LEDs on PDA are both green
- Found: OK and Ready LEDs in the MassLynx Inlet Method Status are both green
- Found: When rebooting PDA, lack of beeping indicates hardware error
- Found: Resetting the IP addresses in the the Waters DHCP Server Configuration does not resolve
ENVIRONMENT
- ACQUITY H-Class System
- ACQUITY PDA Detector
- ACQUITY QDa Detector
- MassLynx 4.1 SCN 947
CAUSE
Antivirus software is interfering with communication between the ACQUITY MS Console and the ACQUITY PDA Detector.
FIX or WORKAROUND
- Remove antivirus software.
- Reboot the system.
ADDITIONAL INFORMATION