Error: Source housing temperature over 90 °C - WKB125760

Article number: 125760

SYMPTOMS

  • MS error: Source housing temperature over 90 °C
  • Source housing temperature shows 110 °C, but it was not heated, not even warm
  • Source heater does not respond to commands
  • Rebooting MS or MassLynx does not help

ENVIRONMENT

  • SQ Detector 2
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-GC
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • Xevo TQ Absolute

CAUSE

Recent cone cleaning caused door not to shut properly.

FIX or WORKAROUND

  1. Open the source enclosure door and verify that the sample cone is correctly installed.
  2. Close the source enclosure door firmly.

ADDITIONAL INFORMATION

 