High Exhaust Pressure errors - WKB12414
Article number: 12414
SYMPTOMS
- "Warning: High exhaust pressure trip" message
- "Warning: Source exhaust pressure too high" message
ENVIRONMENT
- ACQUITY QDa Mass Detector
- ACQUITY TQD system
- TQ Detector
- Xevo TQ Absolute
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
Kink in the exhaust tubing.
FIX or WORKAROUND
- Identify and straighten any kinks in the exhaust tubing.