Skip to main content
Waters

High Exhaust Pressure errors - WKB12414

Article number: 12414

SYMPTOMS

  • "Warning: High exhaust pressure trip" message
  • "Warning: Source exhaust pressure too high" message

ENVIRONMENT

  • ACQUITY QDa Mass Detector
  • ACQUITY TQD system
  • TQ Detector
  • Xevo TQ Absolute
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

Kink in the exhaust tubing.

FIX or WORKAROUND

  1. Identify and straighten any kinks in the exhaust tubing.

ADDITIONAL INFORMATION

id12414, A-SQ, QDA, SQD2, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD

Not able to find a solution? Click here to request help.