Mass Spectrometer does not pump down - WKB12377
Article number: 12377
SYMPTOMS
- Turbo speed does not exceed 50%
- MS does not pump down
ENVIRONMENT
- Xevo TQ
- Xevo TQ-S
- Xevo TQ-XS
- all other MS instruments
CAUSE
Isolation valve O-ring leaks.
FIX or WORKAROUND
Replace the isolation valve O-ring, order code = 700000902 (O-Ring, 6.1 IDx1.6 Kalrez).
(engineering code is 5711294)
ADDITIONAL INFORMATION