Errors in boot sequence: Attempting dhcpcBootInit()....DHCP Initialized; Attempting dhcpcBootBind()......failed dhcpcBootBind(); Can't load boot file - WKB12229
Article number: 12229
SYMPTOMS
EPC boot sequence shows the following:
"Attempting dhcpcBootInit()....DHCP Initialised.
Attempting dhcpcBootBind()......failed dhcpcBootBind().
Can't load boot file!!"
ENVIRONMENT
- TQD
- Xevo TQD
- Xevo TQ MS
- Xevo TQ-S
- SYNAPT G2
- SYNAPT G2 MS/HDMS
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
CAUSE
Problem is related to the WatersFileService or the WatersDHCPService not communicating properly.
FIX or WORKAROUND
- Stop and start both processes ( WatersFileService and Waters DHCP Server ) using the Windows task manager.
C:\Program Files (x86)\Waters Instruments\WatersFileService.exe
C:\Program Files (x86)\Waters Instruments\WDHCPServerSvc.exe
ADDITIONAL INFORMATION