Article number: 12229

SYMPTOMS

EPC boot sequence shows the following:

"Attempting dhcpcBootInit()....DHCP Initialised.

Attempting dhcpcBootBind()......failed dhcpcBootBind().

Can't load boot file!!"

ENVIRONMENT

  • TQD
  • Xevo TQD
  • Xevo TQ MS
  • Xevo TQ-S
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS

CAUSE

Problem is related to the WatersFileService or the WatersDHCPService not communicating properly.

FIX or WORKAROUND

  1. Stop and start both processes ( WatersFileService and WatersDHCPService ) using the Windows task manager.

ADDITIONAL INFORMATION

 

id12229, A-SQ, A-SQ, MALDIG2SI, Q3100MSD, QDA, RADIAN, SQD, SQD2, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOTQ, XEVOTQA, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQGC, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD