Audible gas leak on ACQUITY QDa or RDA - WKB120623
Article number: 120623
SYMPTOMS
- Intermittent, audible gas leak
- Unusually high nitrogen consumption
ENVIRONMENT
- ACQUITY QDa Mass Detector (KAD, KBD)
- BioAccord System
- ACQUITY RDa Detector
CAUSE
The seals in the internal pinch valve assembly have failed.
FIX or WORKAROUND
Replace the internal pinch valve assembly (700011173).
If the pinch valve seals continue to fail regularly due to customer solvent selection consider fitting 700011623 - Pinch Valve Assembly, Internal (Extended Solvent)
ADDITIONAL INFORMATION
Leak can be seen when the pinch valve area is investigated with a bubbling leak checker.