PAL 3 system randomly pauses when using the Xevo TQ-GC - WKB119605
Article number: 119605
SYMPTOMS
- Xevo GC/PAL 3 system pauses randomly—two or three injections run, and then instrument pauses
- CTC Pal status is missing from inlet editor
ENVIRONMENT
- Xevo TQ-GC
- CTC PAL 3
CAUSE
Communication issues
FIX or WORKAROUND
- Shut down MassLynx.
- Power-off CTC PAL 3.
- Reboot the PC.
- Reboot the Xevo GC.
- Power-on CTC PAL.
- Wait 2-3 min for GC and PAL to initialize.
- Launch MassLynx.
- Load the inlet method editor and confirm that both GC and CTC PAL modules are present.
- Launch the Tune page and check Xevo TQ-GC and GC operation.
- Test the system.
ADDITIONAL INFORMATION
id119605, comms, XEVOTQGC