Cone gas cone/sample cone is hard to remove on the Xevo TQ MS, Xevo TQD, Xevo TQ-GC, Xevo TQ-S, Xevo TQ-S cronos, Xevo TQ-S micro or Xevo TQ-XS - WKB116354

Article number: 116354

SYMPTOMS

  • Cone gas cone/sample cone is hard to remove or cannot be removed

ENVIRONMENT

  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-GC
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

The Belleville washers on the top gas cone positioning blocks are stacked together rather than opposing to create a spring.

FIX or WORKAROUND

  1. Close the isolation valve.
  2. Remove the two hex screws that hold the cone positioning blocks in place.
  3. Orient the Belleville washers correctly ") ( ) (" or "( ) ( )"
  4. Reinstall the cone positioning blocks.

ADDITIONAL INFORMATION