Cone gas cone/sample cone is hard to remove on the Xevo TQ MS, Xevo TQD, Xevo TQ-GC, Xevo TQ-S, Xevo TQ-S cronos, Xevo TQ-S micro or Xevo TQ-XS - WKB116354
Article number: 116354
SYMPTOMS
- Cone gas cone/sample cone is hard to remove or cannot be removed
ENVIRONMENT
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-GC
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
The Belleville washers on the top gas cone positioning blocks are stacked together rather than opposing to create a spring.
FIX or WORKAROUND
- Close the isolation valve.
- Remove the two hex screws that hold the cone positioning blocks in place.
- Orient the Belleville washers correctly ") ( ) (" or "( ) ( )"
- Reinstall the cone positioning blocks.