Low desolvation gas flow on Xevo TQ-S - WKB116288
Article number: 116288
SYMPTOMS
- Low desolvation gas flow on Xevo TQ-S or Xevo TQ-XS
- Error appears in MassLynx
- Condition seen after rebooting the instrument
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ-XS
CAUSE
The source pressure test needs to run and pass before the gas flow will turn on.
FIX or WORKAROUND
Run the source pressure test via the tune page drop-down list.
ADDITIONAL INFORMATION