SYMPTOMS

  • Cannot launch "create analysis" or "create analysis method" in UNIFI

ENVIRONMENT

  • Vion IMS QTof

CAUSE

The Security Update for Microsoft Office 2010 (KB4484530) was installed.

 

FIX or WORKAROUND

Uninstall the Security Update for Microsoft Office 2010 (KB4484530).

ADDITIONAL INFORMATION

 