Xevo G2-XS QTof goes into source standby during calibration - WKB107847

Article number: 107847

SYMPTOMS

  • Xevo G2-XS QTof goes into source standby during calibration
  • Calibration does not finish
  • "The desolvation gas flow is too low" message in MS console

ENVIRONMENT

  • Xevo G2-XS QTof
  • MassLynx 
  • ACQUITY UPLC H-Class

CAUSE

Error on LC is causing inlet error-induced shutdown.

FIX or WORKAROUND

  1. Clear error on LC.
  2. If clearing the error is not possible, temporarily disable error-induced shutdown.

Shutdown1.JPG

Shutdown.png

Shutdown2.png

ADDITIONAL INFORMATION